Error "ORA-01691: unable to extend lob segment" appears in the NuGenesis system log files - WKB63258
Article number: 63258
SYMPTOMS
- The following error appears in the NuGenesis LMS server's log file, or in the SDMS Archive Agent/Data Management/Transport Engine log files:
- ORA-01691: unable to extend lob segment schema.[segmentname] by X in tablespace [TablespaceName]
- Depending on the nature of the module reporting the error, users may not be able to log in to NuGenesis and/or the system may not be able to archive files
-
"ORA-1688: unable to extend table" appears in NuGenesis Oracle alert log
ENVIRONMENT
- NuGenesis 9
- NuGenesis 8
CAUSE
The tablespace named in the error message has run out of space and needs another data file.
FIX or WORKAROUND
Contact Waters Technical Support for assistance in resolving errors with the Oracle database for NuGenesis.