An error is reported in wrapper.log for the NuGenesis SDMS Instrument Agents when running as a service - WKB61222
Article number: 61222
SYMPTOMS
- The following error message appears in the wrapper.log file for the SDMS Instrument Agents each time the Waters SDMS Instrument Agents service is started:
- ERROR | wrapper | DATE TIME | The value of property 'wrapper.app.parameter.2', '"ngprojmgr" "manager1" "<Local Servers>" "SERVER"' contains unquoted spaces and will most likely result in an invalid java command line
ENVIRONMENT
- NuGenesis 8 SDMS
- NuGenesis 9 SDMS
CAUSE
The default wrapper.conf file places all of the parameters to the instrument agents on one line.
FIX or WORKAROUND
- Edit the file:
- Drive:\Program Files\Waters\SDMSInstrumentAgents\agents\bin\windows\wrapper.conf
- Look for the following line:
- wrapper.app.parameter.2="ngprojmgr" "manager1" "<Local Servers>" "SERVER"
- Separate the parameters into four lines, as follows:
- wrapper.app.parameter.2="ngprojmgr"
- wrapper.app.parameter.3="manager1"
- wrapper.app.parameter.4="<Local Servers>"
- wrapper.app.parameter.5="win2012-64-mm6"
- Save and close the file.
- Restart the service.
ADDITIONAL INFORMATION
id61222, SDMS, SDMS8, SDMS8NU, SUPISDMS, SUPNG