Error "Out of string space" appears in the NuGenesis Archive Agent log file - WKB60798
Article number: 60798
SYMPTOMS
- The following message appears in the NuGenesis Archive Agent log file:
- Execute Agent. (14): Out of string space.
ENVIRONMENT
- NuGenesis 9 SDMS
- NuGenesis 8 SDMS
CAUSE
One or more of the templates selected within the AA configuration is too large to be processed by the Archive Agent.
FIX or WORKAROUND
- Copy the template lines into a new archive template:
- Open the current archive template for editing.
- Switch to SDMS Administrator and create a new archive template.
- Switch to the current template, select one of the active template lines, and click the Copy button.
- Switch to the new template and click Paste.
- Repeat steps 3 and 4 until all active lines are copied into the new template.
- Save the new archive template.
- Update the AA configuration to use the new template:
- Log in to the SDMS Service Configuration Console (SCC).
- Stop the AA module if it is running.
- In the list of templates, delete the entry for the old template.
- Add the new template to the list.
- Click the Save button and restart the AA module.