Error "[NG14426] Project level Managed store is not found" appears in TE_TEMP folder - WKB60491
Article number: 60491
SYMPTOMS
- Files are .mark_bad in NuGenesis TE_Temp folder
- The following message appears in the TE_TEMP folder:
- ErrorString1=14426:1:3406:[NG14426] Project level Managed store is not found.
ENVIRONMENT
- NuGenesis 9 SDMS
- NuGenesis 8 SDMS
CAUSE
The "Managed Storage" option is set in the project properties, but no Managed Storage Location is configured for the project.
FIX or WORKAROUND
Identify the project:
- Open a .mark_bad file in Notepad.
- Look for the text "DataBaseName=".
- The project name will be next to that text.
Configure the project for managed storage:
- Log in to NuGenesis Administrator.
- Select the project under Master List.
- Right-click the project and click Properties.
- Select the Managed Storage Locations tab.
- Add the Managed Storage device to the project.
- Click OK to close the Properties window.
- Click the Save button. Enter a username, password, and reason for change if prompted.
ADDITIONAL INFORMATION