Error: "TRANSPORT ERROR GETTING VOLUME INFORMATION. Error Category: 1 System Code (53) - The network path was not found" appears in the NuGenesis Archive Agent log file - WKB59013

Article number: 59013

SYMPTOMS

  • No data is archived into NuGenesis SDMS from a particular UNC path \ shared folder
  • Error "GetFSInfofromOcx 10048 Err Desc - TRANSPORT ERROR GETTING VOLUME INFORMATION. Error Category: 1 System Code (53) - The network path was not found" appears in the SDMS Archive Agent activity log file for the template line

ENVIRONMENT

  • NuGenesis 9 SDMS
  • NuGenesis 8 SDMS

CAUSE

The machine named in the UNC path for the template line - or in the "Remote DCOM Server" for Empower/Chromeleon template lines - is down, or for some other reason is no longer reachable on the network.

FIX or WORKAROUND

Start the machine named in the UNC path/Remote DCOM Server and/or fix the network issue.

ADDITIONAL INFORMATION

An example message from the SDMS Archive Agent's log file, NGSDMS80AAActivity.log:

Message........: DATE TIME: Beginning execution of archive request 1.
Message........: DATE TIME: Server: NUG, Project: Empower, Template: EmpServer.
Message........: DATE TIME: GetFSInfofromOcx 10048 Err Desc - TRANSPORT ERROR GETTING VOLUME INFORMATION. Error Category: 1  System Code (53) - The network path was not found.

Message........: DATE TIME: ExecuteTemplateCommand. (0): .
Message........: DATE TIME: An error occurred while performing rollback collection. (91): Object variable or With block variable not set.

Not able to find a solution? Click here to request help.