Error "Access to the registry key 'HKEY LOCAL MACHINE\SOFTWARE\NuGenesis Technologies Corp.\NuGenesis\8.0\EVS\Vendor Registry\Waters Corp.' is denied" appears in the NuGenesis Archive Agent log file - WKB59006
SYMPTOMS
- No Empower projects are archived into NuGenesis
ENVIRONMENT
- NuGenesis 8 SDMS
- NuGenesis 9 SDMS
- NuGenesis 8 Data Adapters
- Empower 3
CAUSE
The Windows account that runs the NuGenesis RPC service is not a member of the local Administrators group of the Empower remote machine.
FIX or WORKAROUND
Add the account to the local Administrators group.
ADDITIONAL INFORMATION
The following messages appear in the Archive Agent's log file, Drive:\Program Files (x86)\NuGenesis 8.0\AA\NGSDMS80AAActivity.log:
"E ","DATE TIME ","5 ","5 ","Access to the registry key 'HKEY_LOCAL_MACHINE\SOFTWARE\NuGenesis Technologies Corp.\NuGenesis\8.0\EVS\Vendor Registry\Waters Corp.' is denied. ","Waters.SDMS.WindowsService ","clsEmp2PreProc ","GetProjectList ","1 ",""
"E ","DATE TIME ","-1 ","-1 ","GetProjectList failed. ","Waters.SDMS.WindowsService ","clsEmp2PreProc ","PerformPreProcess ","1 ",""
"E ","DATE TIME ","-1 ","-1 ","error in Remote PreProcess method ","Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""
"E ","DATE TIME ","-1 ","-1 ","PreProcessor call failed for the remote server: VCL ","Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""