Error "Error encountered while invoking remote EVS Server objects" appears in the NuGenesis Archive Agent log file - WKB58835

Article number: 58835

SYMPTOMS

  • No Empower data is archived into NuGenesis

ENVIRONMENT

  • NuGenesis 9 SDMS
  • NuGenesis 8 SDMS
  • Empower 3

CAUSE

The installed NuGenesis Data Adapters are not compatible with the installed version Empower 3 on the "Remote DCOM Server" machine.

FIX or WORKAROUND

  • Upgrade to the latest available Data Adapters release for your version of NuGenesis, as the latest adapter will likely support more versions of Empower 3 compared to older versions, and it will contain many bug fixes and enhancements
  • If the latest Data Adapter release does not support the installed version of Empower, then consider an upgrade to a more recent version of Empower 3

ADDITIONAL INFORMATION

The following messages appear in the Archive Agent's log file "Drive:\Program Files (x86)\NuGenesis 8.0\AA\NGSDMS80AAActivity.log":

Message........: 2018-11-15 11:30:26: Beginning execution of archive request 1.
Message........: 2018-11-15 11:30:26: Server: SDMSQC, Project: OQFile_Emp_20181114, Template: Empower Archive 81.
Message........: 2018-11-15 11:30:26: Error encountered while invoking remote EVS Server objects. :Could not load file or assembly 'Oracle.DataAccess, Version=2.112.2.0, Culture=neutral, PublicKeyToken=89b483f429c47342' or one of its dependencies. The system cannot find the file specified..
Message........: 2018-11-15 11:30:26: Activitylog LogEntry: [NG14849] Failed to read field info list..
"E ","2018-11-15 11:30:26  ","5 ","5 ","Error encountered while invoking remote EVS Server objects. :Could not load file or assembly 'Oracle.DataAccess, Version=2.112.2.0, Culture=neutral, PublicKeyToken=89b483f429c47342' or one of its dependencies. The system cannot find the file specified.. ","Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""
Message........: 2018-11-15 11:30:26: EMC: PerformPreProcess error.
Message........: 2018-11-15 11:30:26: Archive scan with no files archived for command id 1. {"SDMSQC","OQFile_Emp_20181114","Empower Archive 81"} {"Waters Corp.","Empower-3 Projects","0"}.{Summary : (Total Scanned Files = 0)(Start Time = 15-11-2018 11:30:26)(End Time = 15-11-2018 11:30:26)(Total Time = 0h 0m 0s)}.
Message........: 2018-11-15 11:30:26: Next scheduled runtime for archive request 1 is 2018-11-15 11:31:47.