Error "GetExceptionServicesList: Error in getting Exceptions Services List" appears in the NuGenesis Archive Agent log file - WKB58718
SYMPTOMS
- No Empower data is archived into NuGenesis 8 SR2
ENVIRONMENT
- NuGenesis 8 SDMS
- NuGenesis 8 Data Adapters Release 3 HF2
- Empower 3 SR3
CAUSE
The Exception List was not saved during the File Capture Template Request Line creation.
FIX or WORKAROUND
- Open the File Capture Template, edit the Request Line, and save the Exception List. (You may need to toggle one of the project check boxes in order to enable the "Apply" button.)

ADDITIONAL INFORMATION
The following messages appear in the Archive Agent's log file "Program Files (x86)\NuGenesis 8.0\AA\NGSDMS80AAActivity.log":
Message........: 2019-02-14 09:45:25: Replacing current template , revision 0 with the latest template LempoArchive, revision 1.
Message........: 2019-02-14 09:45:25: 14.2.2019 9:45:25 : GetExceptionServicesList: Error in getting Exceptions Services List. (14599): [NG14599] Cannot find the specified template. It may have been deleted or renamed. Please use a different template..
Fatal error....: 2019-02-14 09:45:25: <- dMacro.LoadExcepsFromToolkit; caught error: Err.Number = -2147221504, Err.Source = "NGArchiveAgentDll", Err.Description = "unexpected retval -1 from pobjTkControl.GetProjectExceptionsList", Err.HelpFile = "", Err.HelpContext = "1000440", Err.LastDllError = "0", Erl = 0.
Message........: 2019-02-14 09:45:25: Startup: An error occured while performing Startup. (5): Invalid procedure call or argument.
Message........: 2019-02-14 09:45:25: Startup of AA Completed.Errors encountered.
Message........: 2019-02-14 09:45:25: <- clsAAProxy.AAExecutControlCommand; retcode: cstFAILURE / cstPROXYFAILURE / gintFailure (-1); byref params: plngErrorCategory = 8, plngErrorCode = 0, plngAuxErrorCode = 0, pstrErrorMessage = "", pstrSourceFile = "", plngLineNum = 0.