Failure in the NuGenesis Transfer App AutoIQ logs for the .ngx/.ngo MIME types - WKB48732
Article number: 48732
SYMPTOMS
- One or more failures similar to the following are reported in the auto IQ log:
- Key: \\HKEY_LOCAL_MACHINE\SOFTWARE\CLASSES\MIME\Database\Content Type\application/nugenesis-sdms
- Name: Extension
- Value: .ngo
- Expected value: .ngo
- Result: Failed
ENVIRONMENT
- NuGenesis 9.0
- NuGenesis 8 SDMS
CAUSE
The "Register OVERLAY MIME Type" program was run and, due to a known defect, registered the wrong program with the file type.
FIX or WORKAROUND
- Fix the "Register OVERLAY" shortcut:
- On the Start screen, right-click the "Register OVERLAY MIME Type" shortcut and click Open file location.
- Right-click the shortcut and click Properties.
- Click into the Target box.
- The rightmost portion of the box reads "|ngx". Change "|ngx" to "|ngo".
- Click Apply and OK.
- Re-establish the NGX and NGO file types:
- Run the "Register OVERLAY MIME Type" shortcut.
- Run the "Register SDMS MIME Type" shortcut.
ADDITIONAL INFORMATION
Defect CRI-75 was filed for the incorrect command line in the "Register OVERLAY" shortcut. Per CRI-75, this defect is fixed in NuGensis 9.1. This shortcut does not exist in NuGenesis 8 SDMS SR1 and earlier systems; however, running the "Register SDMS" shortcut can resolve Auto IQ errors in the SR1 Transfer App.
id48732, SDMS, SDMS8, SDMS8NU, SUPISDMS, SUPNG