Empower project with 0 results is not archived by NuGenesis SDMS when the JCAMP conversion option is active - WKB32112

Article number: 32112

SYMPTOMS

  • Empower projects that do not contain results are not archived into NuGenesis
  • The following messages appear in the NuGenesis Archive Agent activity log:
    • "Value cannot be null.
      Parameter name: o: ","Waters.SDMS.WindowsService ","clsEmp2PreProc ","NGPerformConvTasks ","1 ",""

ENVIRONMENT

  • NuGenesis 8 SDMS
  • NuGenesis 8 Data Adapters R3
  • A file capture template is configured to archive Empower projects with JCAMP conversion

CAUSE

A software defect in the data adapter

FIX or WORKAROUND

Use the Project Exceptions List in the SDMS file capture template to exclude 0-result projects from capture.

ADDITIONAL INFORMATION

Defect reports CRI-38 and CRI-39 were filed for this issue. It was fixed in NuGenesis 9 SDMS Data Adapters R1.