Error "Failed to open Source File. System Error 5 - Access is denied" reported by NuGenesis Data Management - WKB32054
Article number: 32054
SYMPTOMS
- NuGenesis SDMS Data Management reports error in the Activity log file: "<Path & File name> - File was not deleted.NGDM Error 40241 - Failed to open Source File. System Error 5 - Access is denied."
- NuGenesis SDMS Data Management reports error in the Activity log file: "<Path & File name> - file is read-only.File was not deleted.NGDM Error 40245 - Delete File Operation Failed. System Error 5 - Access is denied."
ENVIRONMENT
- NuGenesis 8 SDMS
CAUSE
The NuGenesis DMM module attempts to delete files from an instrument PC that has the Read Only attribute set.
FIX or WORKAROUND
Files must have the Read Only attribute cleared before the NuGenesis DMM module will delete them.
ADDITIONAL INFORMATION