Error when opening an audit trail in NuGenesis LMS: Problems in ExperimentServiceImpl.buildAuditTrail(String, ...) - WKB28070
Article number: 28070
SYMPTOMS
- The following error message appears when opening a document audit trail:
- com.creon.erm.common.SystemException: java.lang.StringIndexOutOfBoundsException: String index out of range: 17
Problems in ExperimentServiceImpl.buildAuditTrail(String, String, boolean, boolean, boolean, boolean, boolean, boolean, boolean, String)
- com.creon.erm.common.SystemException: java.lang.StringIndexOutOfBoundsException: String index out of range: 17
- The document audit trail does not appear
- The audit trail report works for other NuGenesis LMS documents
ENVIRONMENT
- NuGenesis 9 LMS
- NuGenesis 8 LMS
- NuGenesis 8 ELN
CAUSE
An invalid Modification date in the database
FIX or WORKAROUND
Solving this issue requires direct updates to the NuGenesis LMS database tables. Please contact Waters Technical Support for assistance with this issue.