NuGenesis SDMS is not archiving projects from an Empower server located in another time zone - WKB27793
Article number: 27793
SYMPTOMS
-
The Empower and NuGenesis servers are in different time zones
-
Empower project is captured on the first File Capture Scan
-
If the Empower project is modified, the project is not captured again
ENVIRONMENT
- NuGenesis 9 SDMS
- NuGenesis 8 SDMS
CAUSE
Typically, this occurs because the Empower server and the NuGenesis File Capture server are in different time zones. The date comparisons determine if an Empower project is a candidate for archive.
FIX or WORKAROUND
- The Empower server and the NuGenesis File Capture server are required to be in the same time zone.
ADDITIONAL INFORMATION
Defect report CRI-4421 has been filed for this issue.
id27793, SDMS, SDMS8, SDMS8NU, SUPISDMS, SUPNG