SDMS Administrator fails to install when using the NuGenesis 9.3 Bundle installer - WKB258061
Article number: 258061
SYMPTOMS
- NuGenesis SDMS Administrator is selected for installation by the bundle installer
- The installation process appears to finish, and the bundle installer requests a reboot
- After the reboot, SDMS Administrator is not installed
- The installation log file, C:\ProgramData\Waters\NuGenesisBundle\Logs\sdms_adminstrator.log, shows a result code of -3
ENVIRONMENT
- NuGenesis 9.3
CAUSE
Unknown; possibly missing domain information for the NuGenesis RPC service in the Bundle installer setup.
FIX or WORKAROUND
- Restart the Bundle installer and specify a username, password, and domain for the NuGenesis RPC service when prompted by the installer.
- If the installation process still fails, manually install SDMS Administrator from the NuGenesis ISO file.