With NuGenesis 9, mark bad files in TE Temp and Microsoft FTP service stopped - WKB227159
Article number: 227159
SYMPTOMS
- Mark bad files stuck in TE_Temp
- Error in mark bad file: Exception Occured
NGID=110383 [SDMSTransportEngineDetail] Error1=-1 ErrorString1=0:1:571:Exception occured
ENVIRONMENT
- NuGenesis 9, all versions
CAUSE
In services, the Microsoft FTP service is stopped/disabled.
FIX or WORKAROUND
- Open Services.
- Start the "Microsoft FTP Service".
- Right-click on the Microsoft FTP service and go into properties.
- If startup type is disabled, change to "Automatic".
- Open Internet Information Services (IIS).
- Ensure that the FTPs are running.
ADDITIONAL INFORMATION
The Microsoft FTP Service was disabled on the SDMS 9. Ensure that the FTP services on all of the managed store/application servers for SDMS 9 are white-listed to allow the FTP service to be automatically started and in a running state in order for the application to function properly.