"Error while opening CLOB file name" appears in the NuGenesis SDMS system activities project - WKB22397
Article number: 22397
SYMPTOMS
- Transport Engine error in the SDMS_System_Activities Project:
"Error while opening CLOB file name"
ENVIRONMENT
- NuGenesis 8 SDMS
CAUSE
- A reboot of the server or a restart of the NuGenesis RPC service while files are captured
- Temporary drop in the network connection
FIX or WORKAROUND
-
Delete or move all the .mark_bad files away from TE_TEMP into a new folder.
-
Restart the RPC service.
ADDITIONAL INFORMATION
This error message means that all the files in the group that are needed to insert the record in question are not in TE_TEMP.