How to disable LDAP authentication in NuGenesis LMS - WKB20565
Article number: 20565
OBJECTIVE or GOAL
Turn off LDAP authentication in NuGenesis LMS.
ENVIRONMENT
- NuGenesis 8 LMS
- NuGenesis 8 ELN
PROCEDURE
- Log in to the NuGenesis LMS client.
- Open the Administration page > System Configuration. Expand External systems > External Authentication.
- Change the selection on the Authentication drop-down list from "LDAP" to blank (empty).
- Click Apply, and then click OK.
- Close or log out of the LMS client. The LMS authentication system will be used on the next user login.
ADDITIONAL INFORMATION
To disable LDAP authentication with a SQL script:
- Run the VP-ELN_Disable_LDAP_r1.sql script attached.
- This may be the only option if LDAP authentication is active and the LDAP server is down, or if the LDAP configuration is not valid.
Note: A SQL script to enable LDAP authentication is also attached.