How to change the FTP storage path for a NuGenesis Managed Storage location - WKB19658
Article number: 19658
OBJECTIVE or GOAL
Define best practices for moving the SDMS Managed Storage files from one path to another. Often this process is started as part of a NuGenesis upgrade/migration process, or when the storage system backing the FTP site/OSM device is upgraded and given a new path.
ENVIRONMENT
- NuGenesis 8 SDMS
- NuGenesis 9 SDMS
PROCEDURE
- Stop the Microsoft FTP Service on the FTP server
- Use a tool to calculate file hashes for every file in the Managed Storage FTP site
- Use a robust file copy tool, such as RoboCopy, to replicate the current folder structure of the FTP storage path in the new storage path
- Calculate file hashes for every file in the new location
- Compare the file hashes and proceed with the rest of the process only if there are NO discrepancies in the file hashes
- In IIS Manager, select the Managed Storage FTP site, and click the "Basic settings" link
- Set the Physical Path to the UNC path for the new storage location
- Example: \\newServer\SDMSdata
- Click OK
- Restart the Microsoft FTP Service
ADDITIONAL INFORMATION
id19658, SDMS, SDMS8, SDMS8NU, SUPISDMS, SUPNG