Error "Single-row subquery returns more than one row" in NuGenesis Stability pull list - WKB17902

Article number: 17902

SYMPTOMS

  • The following error message appears in NuGenesis Stability when opening the pull list:
    • Single-row subquery returns more than one row

ENVIRONMENT

  • NuGenesis 8 LMS FR1
  • NuGenesis 8 LMS SR2

CAUSE

The Stability database contains some duplicate entries in the following conditions:

  • A user adds a non-pull event to a study protocol for a time interval that is not in the protocol
  • The study protocol is copied to a new study

FIX or WORKAROUND

  1. Look through all of the studies in the NuGenesis Stability database for ones that have a non-pull event in the protocol.
  2. Delete the non-pull event.
  3. Save and close the protocol editor.
  4. Right-click the study and choose Edit Protocol Template.
  5. Click the Intervals tab.
  6. Select the appropriate interval for the non-pull event. Click OK.
  7. Right-click the study and click Edit Protocol.
  8. Select the time interval and add the non-pull event.

ADDITIONAL INFORMATION

This issue stems from a defect in the NuGenesis Stability module. H&A Scientific Inc has registered KBN-761 for this issue. The Waters SCR number is D_031754_06946.