Error: "The NT Service Manager could not be opened to check for the presence of the NuGenesis RPC service" - WKB17888
Article number: 17888
SYMPTOMS
- The following message appears when logging in to the NuGenesis SCC Manager:
- The NT Service Manager could not be opened to check for the presence of the NuGenesis RPC 8.0 Service.
ENVIRONMENT
- NuGenesis 8 SDMS
- NuGenesis 9 SDMS
CAUSE
The NuGenesis Service Configuration Console is running without administrative rights.
FIX or WORKAROUND
- Right-click on the SDMS Service Configuration Console
- Select Run as Administrator
ADDITIONAL INFORMATION