Error "Unexpected error from external database driver (1)" when opening the NuGenesis Connectors mapping tool in LMS - WKB17686
Article number: 17686
SYMPTOMS
- The following error message appears when opening the NuGenesis Connectors mapping tool in the NuGenesis LMS client, or when verifying the credentials for the Connectors Linker service:
"The file containing the error message cannot be loaded. Unexpected error from external database driver (1) . . ."
ENVIRONMENT
- NuGenesis 8 LMS
- NuGenesis Connectors
- Windows Server 2008
- Windows Server 2012
CAUSE
Two patches for Microsoft Windows, KB4041678 and KB4041681, broke the Jet database driver that the Mapping Tool uses to read Excel files.
FIX or WORKAROUND
- Download the older, unpatched version of msexcl40.dll to the client.
- Copy the unpatched file—version 4.0.9756.0—to an arbitrary folder on the client machine, such as C:\JetDB.
- Edit the following registry key to use the path of the unpatched msexcl40.dll file:
-
HKEY_LOCAL_MACHINE\SOFTWARE\Wow6432Node\Microsoft\Jet\4.0\Engines\Excel
Name: Win32
Type: REG_SZ
Data: C:\JetDB\msexcl40.dll
-
-
Close and reopen the LMS client.
ADDITIONAL INFORMATION