Error "invalid username/password" when configuring the NuGenesis Print on Sign Off settings in Empower - WKB16780

Article number: 16780

SYMPTOMS

  • The following error messages appear after clicking the Connect button in the NuGenesis SDMS Archive Properties dialog box in Empower:
    • "[NG14837] invalid username/password; logon denied"
    • "An error occurred while attempting to communicate with the NuGenesis Toolkit."

ENVIRONMENT

  • NuGenesis 8 SDMS
  • Empower 3

CAUSE

The NuGenesis toolkit login key is not active on the Empower machine.

FIX or WORKAROUND

  1. Copy the file NG80ToolkitLoginKey.exe to the Empower machine. It is available on two NuGenesis software part numbers: NuGenesis 8 LMS SR2 Base Media (667005472 revision A) and NuGenesis 8 LMS Power of Two (667005405 revision A). It is also downloadable from this article.
  2. Copy the file into the Drive:\Empower\Bin folder.
  3. Open a command prompt with admin rights.
  4. Run the following command: NG80ToolkitLoginKey.exe ON
  5. Restart any open Empower programs, and then re-enter the SDMS configuration parameters and click the Connect button.

ADDITIONAL INFORMATION

 