Database error: ORA-00600: internal error code, arguments: [pesldl03_MMap: errno 1 errmsg Operation not permitted - WKB123836
Article number: 123836
SYMPTOMS
Error in the alert log in NuGenesis databases:
- ORA-00600: internal error code, arguments: [pesldl03_MMap: errno 1 errmsg Operation not permitted
- ORA-06508: PL/SQL: could not find program unit being called: "ELNPROD.MEASUREORDERUPDATEINDEX"
ENVIRONMENT
- NuGenesis 9.1
- Oracle Database 19c running on Linux
- NuGenesis 9.0
- Oracle Database 12c or 19c running on Linux
CAUSE
Oracle document 1625010.1 describes an issue in the database that can occur when compiling PL/SQL code on Linux databases in 11.1 and later versions. The issue is with a special file in the Linux file system, /dev/shm, when that file is set to read-only.
FIX or WORKAROUND
- Refer to Oracle document 1625010.1.
- The note describes the steps needed to resolve the issue. It is recommended that the database be restarted after making the changes to /dev/shm, even though the Oracle document does not mention it.