Error "Logon failure: unknown user name or bad password" in the NuGenesis Archive Agent log file - WKB10296
Article number: 10296
SYMPTOMS
- The following error message appears in the NuGenesis Archive Agent log file:
- "Error category : 1 System Code (1326) - Logon failure: unknown user name or bad password."
ENVIRONMENT
- NuGenesis 8 SDMS
CAUSE
The username assigned to the NuGenesis RPC service on the NuGenesis AA server is not known to the machine that the AA tried to access.
FIX or WORKAROUND
- Change the NuGenesis RPC service account to one that is recognized by all of the machines to which it will connect. Using a domain account for this purpose is recommended because all machines in the domain will then recognize the account.
- If the Archive Agent will access a machine outside of its own domain, ensure that a trust relationship exists between the domains so that the client machine will accept the domain account.
- If the Archive Agent will connect to a machine that is not joined to a domain, define a local user account on the machine with the same name and password as the domain account.
ADDITIONAL INFORMATION