TargetLynx integrates the wrong peak - WKB91163

Article number: 91163

SYMPTOMS

  • TargetLynx integrates the wrong peak

ENVIRONMENT

  • TargetLynx
  • MassLynx

CAUSE

The process was performed with the minimum peak area = 0, and a peak such as noise having a retention time close to the Predicted Retention Time was recognized.

FIX or WORKAROUND

  1. Click the Edit Method in TargetLynx.
  2. On the Integration Properties screen, click the + mark of Threshold Parameters.
  3. Set the Threshold Absolute Area with an Area value larger than the noise peak, such as 1000.

ADDITIONAL INFORMATION