The shutdown method is set but it does not work well with MassLynx - WKB89146
Article number: 89146
SYMPTOMS
- Although the shutdown method is set, it does not function after the end of the run and flow continues.
ENVIRONMENT
- MassLynx
CAUSE
The Console performs the cleaning with a different method during shutdown.
Also, the settings for unused items, such as Lamp, are included in the shutdown method.
FIX or WORKAROUND
- Create an analytical method that has a unified inlet method configuration, and perform the measurement and cleaning.
- For the shutdown method, delete unnecessary settings (such as the Lamp to off when the lamp is not used) and create the method.
- Perform the measurement with the method created above.
ADDITIONAL INFORMATION
id89146, MLYNX, MLYNXV41, SUPMM