The shutdown method is set but it does not work well with MassLynx - WKB89146

Article number: 89146

SYMPTOMS

  • Although the shutdown method is set, it does not function after the end of the run and flow continues.

ENVIRONMENT

  • MassLynx

CAUSE

The Console performs the cleaning with a different method during shutdown.

Also, the settings for unused items, such as Lamp, are included in the shutdown method.

FIX or WORKAROUND

  1. Create an analytical method that has a unified inlet method configuration, and perform the measurement and cleaning.
  2. For the shutdown method, delete unnecessary settings (such as the Lamp to off when the lamp is not used) and create the method.
  3. Perform the measurement with the method created above.

ADDITIONAL INFORMATION

 

id89146, MLYNX, MLYNXV41, SUPMM