Use Switch2 event when using Divert Valve with MassLynx QDa - WKB86285
Article number: 86285
SYMPTOMS
- Divert Valve not working with MassLynx QDa
ENVIRONMENT
- MassLynx
- QDa
- Divert Valve
CAUSE
Switch1 cannot be selected in Event of MS Method
FIX or WORKAROUND
- Open the MS Method used for the analysis.
- Open Method Event from the Option menu.
- Add the time event of Switch2 to the time for divert valve action.
- If ON is selected for Action, it is sent to the MS; If OFF is selected, it is sent to the waste side.
ADDITIONAL INFORMATION
