IntelliStart error: No calibration file was found for this analysis. The existing calibration will be cleared - WKB80758

Article number: 80758

SYMPTOMS

  • Error appears when performing sample tune and develop method
  • IntelliStart defaults to uncal
  • The correct type of analysis is listed for calibration
  • Mass shifts may be present

ENVIRONMENT

  • MassLynx
  • IntelliStart

CAUSE

Corruption of the calibration file

FIX or WORKAROUND

Create a new resolution and calibration.

ADDITIONAL INFORMATION

Ensure antivirus, firewall and UAC are not interfering as those services can affect the calibration file creation / retrieval.