MassLynx asks for username and password upon launching - WKB79059
Article number: 79059
SYMPTOMS
- MassLynx asks for username and password upon launching
- New installation of software for an LC-only system
- Request for user credentials was unexpected
- Found: Non-regulated laboratory
- Found: MassLynx Security features not required
ENVIRONMENT
- MassLynx
- MassLynx Security
CAUSE
Software was installed with security components activated.
FIX or WORKAROUND
- Remove MassLynx,
- Reinstall MassLynx, selecting neither Full nor Basic Security in the installation wizard.