How to set up a Post Run method in MassLynx - WKB7927
Article number: 7927
OBJECTIVE or GOAL
Set up a washing routine to run immediately after an acquisition.
ENVIRONMENT
- MassLynx 4.1 SCN 916
- SYNAPT G2-Si
- ACQUITY UPLC H-Class
PROCEDURE
- Create the LC method in the current MassLynx project using the Inlet Method Editor.
- Save the method (best to use a differentiating title like "PostRun" or "Wash").
- Because there are no settings in the LC method to define the flow path, open the MS method and check the Method Events.
- If necessary, adjust the Method Events (immediately before the end of the run time) to set the flow path.
- Save the MS Method.
- Customize the MassLynx Samplelist table to include the INLET_POSTRUN column.
- Browse for the post-acquisition LC method in the INLET_POSTRUN column of the Samplelist.
- Save the Samplelist and test.
ADDITIONAL INFORMATION
id7927, MLYNX, MLYNXV41, SCN916, SUPMM