How do you configure a Mass Spectrometer to start the acquisition by contact closure when the inlet is not supported by MassLynx? - WKB7772
Article number: 7772
OBJECTIVE or GOAL
Trigger a Waters MS to start an acquisition via contact closure in MassLynx.
ENVIRONMENT
- MassLynx
- Xevo TQ Absolute
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-GC
- Xevo TQ-S
- Xevo TQ-S cronos
- Xevo TQ-S micro
- Xevo TQ-XS
PROCEDURE
- Open the MassLynx inlet method.
- Go to Tools > Instrument Configuration.
- Click the Events and Triggering button.
- Click NEXT on the first page of the wizard.
- On the Choose Events page of the wizard, select Event In 1.
- At the end of the wizard, click Finish. Click Finish to close the instrument configuration.
- Connect a contact closure cable from Inject Start out of the autosampler to Event In 1 on the back of the MS.
- On the sample list, in the Inlet Method column, type the word Default.
ADDITIONAL INFORMATION
id7772, A-SQ, A-SQ, CYCIMS, MALDIG2SI, MLYNX, MLYNXV41, MRT, Q3100MSD, QDA, QTOF-PREM, RADIAN, SQD, SQD2, SUPMM, SYNAPT, SYNAPTMS, SYNAPTXS, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, TQD, TQDIVD, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSQT, XEVOG2XSTF, XEVOG3QTOF, XEVOQTOF, XEVOTQ, XEVOTQA, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQGC, XEVOTQIVD, XEVOTQS, XEVOTQSCRO, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD