DynamX processing error "Apex3D64.exe aborted with unexpected error code: -1" - WKB76762

Article number: 76762

SYMPTOMS

  • Attempt to process data through DynamX fails with error "Apex3D64.exe aborted with unexpected error code: -1" for each time point
  • View log on the DynamX software shows the last entry as follows:
    "Survey Low Energy to calculate peak width
    Chrom peak width
    Error: Failed to calculate chrom peak width
    Exiting"

ENVIRONMENT

  • DynamX HDX data analysis software
  • HDX
  • HDX-2

CAUSE

Review of RAW data file reveals that this is a blank injection that contains no peptides. DynamX is searching for a peptide peak to calculate peak width.

FIX or WORKAROUND

Inject a sample with actual peptides present for calculation.