DynamX processing error "Apex3D64.exe aborted with unexpected error code: -1" - WKB76762
Article number: 76762
SYMPTOMS
- Attempt to process data through DynamX fails with error "Apex3D64.exe aborted with unexpected error code: -1" for each time point
- View log on the DynamX software shows the last entry as follows:
"Survey Low Energy to calculate peak width
Chrom peak width
Error: Failed to calculate chrom peak width
Exiting"
ENVIRONMENT
- DynamX HDX data analysis software
- HDX
- HDX-2
CAUSE
Review of RAW data file reveals that this is a blank injection that contains no peptides. DynamX is searching for a peptide peak to calculate peak width.
FIX or WORKAROUND
Inject a sample with actual peptides present for calculation.
ADDITIONAL INFORMATION
