Can I use Ethernet control for my Alliance with MassLynx? - WKB6868

Article number: 6868

ENVIRONMENT

  • 2695 Separations Module
  • e2695 Separations Module
  • 2695D Dissolution Separations Module
  • 2795 Separations Module
  • Alliance iS HPLC System
  • Alliance HPLC System
  • IEEE
  • Ethernet
  • MassLynx

ANSWER

No. MassLynx does not support control of any Alliance inlet instrument with ethernet communications.

2695 / 2795 Separations Modules must use a GPIB card / IEEE comms / National Instruments control software when controlled by MassLynx.

e2695 Separations Module is supported by GPIB on models with an IEEE connector but later models only have an ethernet port and cannot be controlled by MassLynx.

Cut over serial numbers for eCPUs without the IEEE port are: F25SM7165A (186269505) & E25SM4154A (186269506)

Alliance iS HPLC System is not supported in MassLynx

ADDITIONAL INFORMATION

Article wkb1277 may be useful: How to configure an Alliance and UV detector on a MassLynx system