2998 PDA "not connected" error in MassLynx - WKB68230
Article number: 68230
SYMPTOMS
- 2998 PDA is configured in MassLynx
- Inlet editor OK and ready lights both green
- Additional Status tab shows 2998 "not connected"
ENVIRONMENT
- 2998 Photodiode Array Detector
- MassLynx
CAUSE
Unknown
FIX or WORKAROUND
- Close MassLynx and associated windows.
- Run StopMassLynxProcesses.
- Open Waters DHCP Server Configuration from C:/Program Files (x86) / Waters Instruments folder.
- Remove the 2998 from the DHCP server.
- Power cycle the 2998 PDA.
- Ensure that the 2998 PDA repopulates in the DHCP server screen.
- Start MassLynx and confirm that communications are now OK, including in the Additional Status tab.
ADDITIONAL INFORMATION
id68230, 2998, 2998B, 2998SFC, A-10PDA, comms, MLYNX, MLYNXV41, SUPMM