QuanLynx and TargetLynx data processing takes 3 minutes to process a 50-minute, full scan Tof data file for integration only - WKB6840
Article number: 6840
SYMPTOMS
- TargetLynx processing taking 3 minutes to process a 50-minute Tof data file
- Integration only
- MS acquisition method: m/z range 50 to 500 amu
- Data file (*.raw) size is approximately 5 GB
ENVIRONMENT
- Xevo G2 XS Tof
- MassLynx 4.1 SCN 939
- TargetLynx
- QuanLynx
CAUSE
The data processing is slow due to the size of the data files. This is expected behavior
FIX or WORKAROUND
- Tof data should not process with QuanLynx. It should only process with TargetLynx.
- Data should process off-line with the desktop version of MassLynx/TargetLynx.
- Data processing is slow due to the size of data files.
ADDITIONAL INFORMATION