SYMPTOMS

  • While installing software, error message appears: "Append Error: Could not find file Flavormap.properties" (see screenshot)

error.JPG

ENVIRONMENT

  • MassLynx 4.2 SCN 985

CAUSE

 Interference from third-party IT/security software

FIX or WORKAROUND

Remove the unsupported software.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.

  • Was this article helpful?