How to update an existing user list in OAToolkit and preserve existing information - WKB60795
Article number: 60795
OBJECTIVE or GOAL
Add information to an existing user list in OAToolkit while preserving passwords and user information.
ENVIRONMENT
- OAToolkit
- OALogin
- MassLynx
PROCEDURE
To import and export user names and password changes in OAToolkit:
- Save your current configuration in OAToolkit (save icon).
- We recommend that you create a copy for backup purposes. In File Manager, copy and save the C:\OAToolkit\config\OALPDPConfig.xml file.
- In OAToolkit, select Users > Import Users…
- In Import Method, select OALogin PC and browse to the appropriate PC and OALogin.pro.
- In the User List, select either Create New List or Add to List, as appropriate.
- Click OK. (The OAToolkit loads with a fresh list containing only the users from the OALogin terminal selected.)
- Make any modifications to the user list (for example, add users, change passwords, or force password resets).
- In OAToolkit, select Terminals > Upload.
- Browse to the OALogin PC, and then select Upload to Selected OR
Users > Export Users > OALPDPConfig.xml File. - The new, modified list of users is available on the OALogin PC.