Skip to main content
Waters

The shutdown method for the ACQUITY TQD did not complete the intended tasks after rebooting the MassLynx PC - WKB6013

Article number: 6013

SYMPTOMS

  • Reboot MassLynx PC after communication issue
  • Shutdown method did not complete intended tasks
  • Shutdown tasks were not listed in the shutdown method

ENVIRONMENT

  • MassLynx
  • ACQUITY TQD

CAUSE

Ran the wrong shutdown file

FIX or WORKAROUND

  1. Select Edit Shutdown or Startup.
  2. Confirm the correct file is open, or select File > Open, then browse to the correct file.
  3. Click Auto Control Tasks to confirm the tasks are listed.
  4. Click Shutdown.
  5. Click Enable shutdown after batch.
  6. Select File > Save.

ADDITIONAL INFORMATION

 

comms

Not able to find a solution? Click here to request help.