In the TargetLynx result file, the Chromatogram and Calibration panes cannot be resized - WKB58734

Article number: 58734

SYMPTOMS

ENVIRONMENT

  • MassLynx 4.1
  • TargetLynx XS
  • MassLynx 4.1
  • ACQUITY SQD system
  • SQ Detector
  • ACQUITY TQD system
  • TQ Detector
  • SQ Detector 2
  • Xevo TQ Absolute 
  • Xevo TQ MS
  • Xevo TQD
  • Xevo TQ-S
  • Xevo TQ-S cronos
  • Xevo TQ-S micro
  • Xevo TQ-XS

CAUSE

Unknown

FIX or WORKAROUND

  1. Close MassLynx.
  2. Browse to C:\MassLynx.
  3. Find the TargetLynx.~ql file and rename it to TargetLynx_old.~ql.
  4. Go to the Windows Start menu.
  5. Type in Regedit in the search box.
  6. In the Registry Editor window, browse to this location in the far left-hand pane: Computer > HKey_LOCAL_Machine > Software > Wow6432node > Micromass > TargetLynx
  7. Delete that TargetLynx folder.
  8. In the Registry Editor window, browse to this location in the far left-hand pane:  Computer > HKey_Current_User > Software > Micromass > TargetLynx XS
  9. Delete that TargetLynx XS folder.
  10. Exit the Registry Editor and restart the computer.

ADDITIONAL INFORMATION

For steps 8 and 9, if the system has TargetLynx instead of TargetLynx XS, the folder is just called TargetLynx.