Skip to main content
Waters

MS goes to Standby mode when running IntelliStart - WKB55201

Article number: 55201

SYMPTOMS

  • Instrument goes to Standby mode when using Sample Tune and Method Development
  • The inlet method OK indicator light is red

ENVIRONMENT

  • Xevo TQ MS
  • Xevo TQD
  • Xevo TQ-S
  • Xevo TQ-S micro
  • Xevo TQ-XS

CAUSE

The inlet is not ready, and the shutdown method is configured to shut down on an "Inlet Fatal Error".

FIX or WORKAROUND

  1. Open the shutdown settings.
  2. Click the Configure Error Shutdowns button.

clipboard_e1610d9e1347b30f8f5a24ef839a3fbf8.png

  1. Clear the check box next to "Inlet fatal error".

clipboard_ed827a000a5046e840494ed1f293bd6d4.png

4.  Reset the "LC communications" on the inlet method.

5.  Confirm that the inlet method OK indicator light is green.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.