MassLynx goes straight to "acquiring", without triggering the inlet to inject - WKB53375
Article number: 53375
SYMPTOMS
- When starting a sample list run in MassLynx, the MS starts acquiring immediately
- The inlet does not perform injection sequence
- Inlet Events and Triggering are configured correctly; Event in 1 is selected
- LC systems running in standalone inject correctly
ENVIRONMENT
- MassLynx
- Xevo TQ Absolute
- Xevo TQ-S
- Xevo TQ XS
- XEVO G2 QTOF
- XEVO G2 XS
CAUSE
Faulty User Event PCB on the back of the mass spectrometer.
e.g. one system had Event In 2 wired to Event In 1, and Event In 1 wired to event In 2.
FIX or WORKAROUND
Fix - Replace the User Event PCB.
For Xevo TQ Absolute: User Event PCB Part Number: 700004382 (USER EVENT PCB ASSY)
Workaround for incorrectly wired PCB - Select Event in 2 in Inlet Configuration Events and Triggering.
ADDITIONAL INFORMATION