How do I remove the ACQUITY UPLC PDA Detector from the Inlet Configuration in MassLynx - WKB53515

Article number: 53515

OBJECTIVE or GOAL

To remove the ACQUITY UPLC PDA Detector from the Inlet Configuration in MassLynx.

ENVIRONMENT

  • MassLynx 4.1, 4.2
  • ACQUITY UPLC PDA Detector

PROCEDURE

  1. Open the Inlet Method Editor in MassLynx.
  2. Go to Tools > Instrument Configuration.
  3. Click Configure.
  4. Follow the Configuration Wizard to the Detector page.
  5. Highlight the ACQUITY UPLC PDA Detector in the list of Configured Detectors (right-hand panel).
  6. Click the left pointing arrow to remove the detector from the configuration.
  7. Click Next, and then Finish (twice).
  8. Close MassLynx and the ACQUITY MS Console.
  9. Run StopAcquityProcesses.cmd.
  10. Open MassLynx and the ACQUITY MS Console.