No analog data recorded in MassLynx during acquisition - WKB52948
Article number: 52948
SYMPTOMS
- No analog data stored; no data points created
- MS data is recorded
- Detector records non-storage data in ACQUITY Console
ENVIRONMENT
- MassLynx
- ACQUITY UPLC ELS Detector
CAUSE
MassLynx doesn't receive the analog data from the detector to store in a data file.
FIX or WORKAROUND
- Go to Tools > Instrument Configurations.
- Configure the system as it is, with an optical detector.
- In the last window, search for instruments.
- Select the correct serial number for the ELSD (or other optical detector) in the drop-down list for the corresponding detector type.