When creating new users in the MassLynx Security Manager, the list of domain users does not show in the username pull-down list - WKB52974
Article number: 52974
SYMPTOMS
- No domain users shown in the username pull-down list
- Software recently reloaded
- Only two user accounts are available in the MassLynx Security Manager: Administrator and Micromass
ENVIRONMENT
- MassLynx
- MassLynx Full Security
- MassLynx Security Manager
CAUSE
The Local Administrator user account on the PC does not have the required permissions to view the list of usernames on the domain.
FIX or WORKAROUND
- Log in to MassLynx Security Manager with the default Administrator user account.
- Go to Create > New User.
- Specify the corporate domain username for the user who will be a MassLynx administrator and have administrative privileges on that domain.
- Specify the Full Name and Description as required.
- Select the correct domain.
- Set the Group Membership for Regulated mode to Administrators.
- Click OK to create the user.
- Go to File > Save.
- Close the MassLynx Security Manager and click OK or YES to the message regarding verifying the user on the domain.
- Launch the MassLynx Security Manager and log in using the Administrator account created in Steps 2 through 8.
- Go to Create > New User.
- Select the domain and pull down the list of users available on that domain.
ADDITIONAL INFORMATION