No concentration results reported in TargetLynx - WKB51128

Article number: 51128

SYMPTOMS

  • No concentration results reported in TargetLynx
  • Peaks are properly integrated
  • Found: Response value reported a 0.00 for all peaks
  • Found: Response type set to "Internal (relative - use internal standard)"
  • Found: Concentration of Standard: Level set to "Fixed", with Concentration of Standard set to 0.00 for the Internal Standard compounds

ENVIRONMENT

  • MassLynx
  • TargetLynx

CAUSE

Concentration of Standard for the Internal Standard compounds is set to 0.00.

FIX or WORKAROUND

  1. Set the Concentration of Standard for the Internal Standard compounds to either the exact concentration or the value 1.00.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.