TargetLynx is not saving peak modifications - WKB509
Article number: 509
SYMPTOMS
- Peak not integrated by method
- Add peak: Peak of interest is integrated
- Save Peak Modification: Peak disappears (no longer integrated)
ENVIRONMENT
- MassLynx 4.1
- TargetLynx
- QuanLynx
CAUSE
Peak height or peak area is below the threshold set in the integration parameters of the method.
FIX or WORKAROUND
- Disable or decrease the Relative Height and Relative Area settings under Threshold Parameters.