What are the parameters for the ANSI-48Vial2mLHolder plate in MassLynx? - WKB50865
Article number: 50865
ENVIRONMENT
- MassLynx
ANSWER
The plate parameters are as follows (also see attached image):
Rows - #Vials=6, Vial Spacing=140
Columns - #Vials=8, Vial Spacing=140
Vial - Diameter=120, Depth=320
Offsets - Column=0, Row=0
Origin - Top Left
Grid - Horizontal=1, Vertical=A, Referencing=XY, Priority=Horizontal First
Plate Size - X=1280, Y=860, Z=383
Top Left Vial Offset - X=150, Y=83
ADDITIONAL INFORMATION
id50865, MLYNX, MLYNXV41, SUPMM, UPBINARY