Skip to main content
Waters

How do you change the needle depth for an ACQUITY Sample Manager in MassLynx? - WKB50312

Article number: 50312

OBJECTIVE or GOAL

Adjust the needle depth on an ACQUITY Sample Manager.

ENVIRONMENT

  • ACQUITY UPLC
  • ACQUITY Sample Manager
  • ACQUITY UPLC I-Class
  • MassLynx

PROCEDURE

  1. Open the Inlet Editor in MassLynx.
  2. Under the Autosampler, click the Advanced button.

clipboard_ece16a489b41a7cd8786f936fbdba1eae.png

  1. Check the box next to "Needle Placement (from bottom)".
  2. Enter a new value (default for the SM-FL is 2 mm, and SM-FTN is 4mm).

clipboard_e9ab740a1e4be4e21ed48b8651d700811.png

ADDITIONAL INFORMATION

Note the minimum recommended needle placement for the different types of vials.

clipboard_e2c7e1eb42b7ef74d7991a44071a2de2a.png

id50312, MLYNX, MLYNXV41, SUPMM, UPBINARY, UPLCCHIVD, UPSMPMGR

Not able to find a solution? Click here to request help.