How to create a new reference file for IntelliStart to use - WKB49211
Article number: 49211
OBJECTIVE or GOAL
Create a new calibration reference file for use within IntelliStart.
ENVIRONMENT
- MassLynx
- SYNAPT G2-Si
- IntelliStart
PROCEDURE
- With MassLynx closed, open C:\MassLynx\Ref and copy in or create a new reference file.
- If creating a new reference file from scratch or using an existing reference file from a previous installation, ensure that the layout/syntax matches that of the current installation's example reference files (see Additional Information below).
- Once the reference file has been saved, browse to the C:\MassLynx\Intellistart\Configuration\ReferenceFile.xml.
- Make a copy of this XML file as a backup, and then right-click and open it with Notepad.
- Insert the new entry between the lines <ReferenceFiles> and </ReferenceFiles> (see example below).
- Ensure that the format matches the existing entries. (Note that the full path must be given and that it is case-sensitive.)
- When edits are complete, save the file.
- Exit Notepad.
- Open MassLynx and verify that IntelliStart shows the new file.
ADDITIONAL INFORMATION
With respect to a reference file layout, here are some important notes.
a - Any line that starts with a semicolon will not be read as a reference mass.
b - The line ;Parent=xxx.xxxx specifies the parent mass if the reference file is to be used in MSMS.
c - The line ;CE=xx specifies the collision energy that IntelliStart will use by default with the reference file. If fragmentation is required, this should be set to an appropriate value.
d - To the right of every reference file is a second value signifying the relative intensity (%) of each reference peak when compared with the reference peak that is most intense (labeled 100).