No communication with Quattro Premier XE - WKB29232
Article number: 29232
SYMPTOMS
- EPC does not boot
- Pinging EPC results in sporadic response
ENVIRONMENT
- Quattro Premier Xe
- Windows 7
CAUSE
Micromass administrator account is not set up on host PC.
FIX or WORKAROUND
Set up Micromass administrator account on host PC.
ADDITIONAL INFORMATION
See KCS article 6468 for information on how to configure a Windows 7 PC to communicate with MassLynx.
id29232, comms, MLYNX, MLYNXV41, QUATPREMXE, QUPREMIER, SUPMM